OneStim™ CRM

The OneStim-CRM™ provides, in one device, the key functionality recommended for implanting conduction system pacing leads according to the 2023 EHRA clinical consensus statement on conduction system pacing implantation [1], namely:

  1. Voltage controlled pacing
  2. Continuous impedance measurements
  3. Current of Injury display, and
  4. 12 Lead ECG display and measurement.
Features
  • Continuous impedance with graph
  • Full 12 ECG display
  • Trigger screen with automatic measurements – RWPT, V6-V1 Int
  • Paced extras
  • Template comparison
  • Review screen with multiple callipers
https://micropaceep.com/wp-content/uploads/2025/11/Capture.png
OneStim™ CRM Specifications
Patient Interface
  • 2 intra-cardiac channels
  • 12-13 lead surface ECG
  • 1 independent emergency pacing channel
Programmable Stimulation
  • Pace, A-V Delay, Nodal ERP, R-synced Sx, Threshold
  • 0.1-25V (50V in esophageal mode)
  • 0.1-25mA
ECG Analysis & Recording
  • Up to 90s buffer of saved ECG for review at sweep speed up to 400mm/s
  • Time and amplitude measurement callipers (±1ms, ±0.1mV)
  • Triggered sweep
  • Template comparison
  • Save printable reports
Training Videos

Reference: [1] Burri, H., Jastrzebski, M., Cano, Ó., Čurila, K., de Pooter, J., Huang, W., Israel, C., Joza, J., Romero, J., Vernooy, K., Vijayaraman, P., Whinnett, Z., & Zanon, F. (2023). EHRA clinical consensus statement on conduction system pacing implantation: endorsed by the Asia Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS), and Latin American Heart Rhythm Society (LAHRS). Europace, 25(4), 1208–1236.